Models 200EH/200EM
Chemiluminescene High &
Medium Range NO/NO2/NOX
Analyzers
The Models 200EH and 200EM use the proven chemiluminescence measurement
principle, coupled with state-of-the-art microprocessor technology for
monitoring high and medium levels of nitrogen oxides. User-selectable
analog output ranges and a linear response over the entire measurement
range make them ideal for a
wide variety of applications, including extractive and dilution CEM,
stack testing, and process control.
A choice of NO2 converters handles tough CEM and
stack testing as well
as combustion turbine startup and continuous operation applications.
Selectable measurement modes (NOX only, NO only,
NO/NOX switching),
auto ranging (single range, dual range, auto-range, independent ranges,
remote range control) enables to match the Models 200EH or 200EM to
your needs.
Modular instrument design offers top-mounted, quick access to all
subassemblies and hinged front and rear panels to simplify module
replacement and maintenance. A standard permeation dryer on the ozone
generator provides dry air and excess ozone is removed by catalytic
reaction, both eliminating the need for expendables.
All instruments in the Teledyne-API Model “E”-
Series include built-in
data acquisition capability with internal memory. This allows logging
of multiple parameters in different predefined or customized data
channels including averaged or instantaneous concentration values,
calibration data, and operating parameters such as pressures and flow
rates.
Stored data are easily retrieved through the RS-232 port via APICOM or
from the front panel, allowing predictive diagnostics and enhanced data
analysis by tracking parameter trends. The Models 200EH and 200EM
combine rugged construction, light weight, ease of use,
powerful diagnostics and outstanding performance for high and medium
range applications.
Features:
- 200EH: 0-5 ppm to 0-5000 ppm, user selectable
- 200EM: 0-1 to 0-200 ppm, user selectable
- Independent ranges for NO, NO2, NOX
- Auto ranging and remote range selection
- Microprocessor controlled for versatility
- NOX-only or NO-only modes
- Multi-tasking software allows viewing of test variables
while operating
- Continuous self checking with alarms
- Permeation drier on ozone generator
- Dual bi-directional RS-232 ports for remote operation
(optional Ethernet or RS-485)
- Digital status outputs provide instrument condition
- Adaptive signal filtering optimizes response time
- Temperature & pressure compensation, automatic
zero correction
- Converter efficiency correction software
- Minimum CO2 and H2O
interference
- Catalytic ozone scrubber
- Internal data logging with 1 min to 365 day multiple
averages (1 million records)
| Specifications: |
| Ranges: |
200EH: 0-5 ppm
to 0-5000 ppm full scale, user selectable; independent NO, NO2,
NOX ranges and auto-ranging supported
200EM: 0-1 ppm to 0-200 ppm |
| Units: |
ppm, mg/m3 |
| Zero noise: |
< 20 ppb
(RMS) |
| Span noise: |
< 0.2% of
reading above 20 ppm (RMS) |
| Lower Detectable
Limit (LDL): |
< 40 ppb
(RMS) |
| Zero drift: |
< 20
ppb/24 hours |
| Span drift: |
< 1%
reading / 24 hours, < 1% reading / 7 days |
| Lag time: |
20 seconds
switching mode; < 6 seconds NO or NOX
only mode |
| Rise and Fall
Time: |
< 60
seconds to 95% (switching); 5 seconds NO only; 10 seconds NOX
only |
| Linearity: |
1% of full scale |
| Precision: |
0.5% of reading |
| Sample Flow Rate: |
250 cm3/min
(standard); 500 cm3/min (optional) |
| Operating
Temperature Range: |
5 - 40° C |
| Dimensions
(HxWxD): |
7”
(178 mm) x 17” (432 mm) x 23.5” (597 mm) |
| Weight: |
Analyzer - 44
lbs (20 kg)
External pump - 15 lbs (7 kg) |
| Power: |
100V 50/60 Hz,
115V 60Hz, 220V 50/60Hz, 230V 50Hz, 240V 50Hz, 250 Watts (analyzer),
250 Watts (pump) |
| Analog outputs: |
10V, 5V, 1V,
0.1V, selectable |
| Recorder Offset: |
±10% |
| Serial Outputs: |
Serial Port 1:
RS-232 (DB-9M)
Serial Port 2: standard RS-232 or optional RS-485 (DB-9F) |
| Status (Digital): |
8 outputs, 6
inputs (opto-isolated), 6 alarm outputs (optional) |
| Current Output: |
0-20 mA or 4-20
mA isolated outputs (optional) |
| Approvals: |
CE |
|
|